This text is a supportive document to our "Response to Miller's Critique on SEE++" and provides a technical evaluation of Orbit™ 1.8 original Macintosh source code as well as a comparison between Orbit™ 1.8 software and Orbit™ model in SEE++.
There is no doubt that the biomechanical model contained in Orbit™ 1.8 is scientifically well supported, established and realistically predicts eye motility for normal "healthy" eyes and a large set of simulations of pathologies. But it is also a fact that the Orbit™ 1.8 biomechanical model has not been updated since 1999, certainly an eternity for a computer program. Moreover, recent physiological findings like "Active Pulleys" are not included in Orbit™ 1.8.
Orbit™ 1.8 was implemented for the Macintosh "Old World" generation (Macintosh II series, Quadra/Centris, PowerPC, G3 etc.), only runs in emulation mode on Mac OS X and cannot run on the new Intel-based Macs. Moreover, the development environment and class library that was used to implement Orbit™ does not exist anymore. This unfortunately must lead to the conclusion that the Orbit™ 1.8 computer program is hopelessly outdated.
We see SEE++ as replacement or extension to Orbit™ with more clinical relevance while providing essential functionality similar to what Orbit™ offered. The SEE++ software differentiates between biomechanical models and user interfaces and therefore provides an open, flexible and portable basis for further development. Additionally, the "SEE-KID" and "SEE-KID Active Pulley" models have been developed in order to incorporate new anatomic and physiological findings from basic research. Compared to Orbit™, these models also use a different mathematical approach for numerical optimization in order to more reliably solve non-linear problems, a special and important task when simulating the statics of mechanical systems like the human eye.
When looking at the source code of Orbit™, an experienced Software-Engineer will get desperate and tear one's hair. Code analysis of the Orbit™ mathematical model core reveals a total of 5123 lines of code, 2672 being comments only, which are mainly uncommented variants of different calculations. This means that more than 50% of the total source code is commented out and meaningless. Software quality finds, amongst other things, to a major part its roots in well structured and exactly designed implementation. The implementation of Orbit™ could at best serve as deterrent example for "how to fail on this topic". It is no matter of course that highly qualified researchers transform their ideas and breaking research results into well-formed maintainable software applications. On the other hand, we do not claim to carry out basic medical or physiological research, however, we understand our work as interdisciplinary applied research that follows basic research as well as clinical practice, creating technically and socially useful solutions.
When it comes to scientific computing, computer programs are used to construct mathematical models and solution techniques to analyze and solve scientific problems. In practical use, this is typically the application of computer simulation and other forms of computation to problems in various scientific disciplines. Unfortunately, most problems in these fields cannot be solved analytically and numerical methods need to be applied to obtain approximations to a solution. Orbit™ as well as SEE++ apply numerical methods to find mechanical force equilibriums for eye positions and innervations. Building on that, both computer programs simulate binocular functions and simulate the clinical Hess-Lancaster test.
When using numerical computing methods, floating point arithmetic is a famous source for computational errors. Once errors are generated, they propagate through calculations and some algorithms amplify these errors which then causes numerical instability. Due to its implementation, Orbit™ 1.8 is vulnerable to such errors.
We would like to support our statements with some of the most obvious implementation problems that we found in the Orbit™ 1.8 source code:
- There are different implementations for the calculation of Listing's Torsion:
There is a function "myListing" defined in the "sqlibrary.cp" file that uses two constants "Degrees" and "Radians" for conversion of degrees to radians and back. These constants are defined as follows:
const extended Radians = 180. / 3.14159;
const extended Degrees = 3.14159 / 180.;
In other cases, there are definitions of the constant PI with higher precision in "squint.h" and "sqnt.h" that are used for conversions of degrees to radians and vice versa:
#define PI 3.141592653589793238462643383
#define DEG_PER_RAD (180.0 / ORBIT_PI)
#define RAD_PER_DEG (ORBIT_PI / 180.0)
It turns out that Orbit™ uses two functions "myListing" and "_listing" to calculate Listing's Torsion, "myListing" for angles in degrees and "_listing" for radians. However, sometimes "myListing" is used which results in low precision calculation of degrees to radians conversion, in other situations "_listing" is directly called with "DEG_PER_RAD" and "RAD_PER_DEG" precalculated high precision radian angles.
In "Fascia.cp", the difference between two Listing's Torsion values, one in low and one in high precision is calculated, which leads to the inconsistency that for exactly the same eye position, this difference is not 0.
One could now argue that these differences are marginal, but in connection with an iterative optimization, these inconsistencies cumulate and result in different model predictions. Deciding whether a solution is right or wrong, compared to other models or calculations becomes a challenging task in this situation.
Unfortunately, if this "bug" is corrected, solutions cannot be compared to the original Macintosh version of Orbit™ 1.8 due to different model predictions that are displayed in the mechanical stateviewer dialog of Orbit™.
- Orbit™ uses float precision numbers for finding 3D eye positions but extended precision numbers anywhere else:
For calculating 3D eye positions from 2D eye positions when the fixing eye is pathologic, Orbit™ uses a slightly modified implementation of the Newton minimization algorithm taken from the book "Numerical Recipes in C". The default implementation uses for all numbers float precision and so does Orbit™. However, all other model code in Orbit™ uses the "extended" floating point type. The "extended" type is a non-standard Macintosh specific data type, which offers more precision in floating point calculations than the "double" type. On Intel platforms (also on the new Intel-Macs) no such data type exists due to the different processor architecture.
extended g_pos[3];
static void POS_POS(int n, float x[], float f[]){
...
...
...
extended rot_deg[4]; /* theta, phi, psi, cyclo */
rot_deg[0] = x[1];
rot_deg[1] = x[2];
rot_deg[2] = myListing(x[1], x[2], psi0);
// see (1) for problems with myListing
...
...
...
f[1] = rot_deg[0]; // implicit extended to float conversion
f[2] = rot_deg[1]; // possible loss of significant digits
g_tor = rot_deg[2];
f[1] -= g_pos[0]; // implicit extended to float conversion
f[2] -= g_pos[1]; // possible loss of significant digits
} /*POS_POS*/
void find_inner_eye_positions(...) {
...
...
...
int n = 2;
float x[3], x_int[3];
x[1] = g_pos[0]; // implicit extended to float conversion
x[2] = g_pos[1]; // possible loss of significant digits
...
...
...
int check = 0;
newt(x, n, &check, POS_POS);
...
...
...
} /*find_inner_eye_positions*/
- Orbit™ passes most parameters internally via strings:
To our big surprise, we discovered that Orbit™ almost randomly uses format strings to cut or round intermediate calculation results. Here are some examples:
//Definitions from squint.h
#define PR_POS_L "%20.15lf %20.15lf %20.15lf %20.15lf %20.15lf %20.15lf\n"
#define PR_EXC_L "%20.15lf %20.15lf %20.15lf %20.15lf %20.15lf %20.15lf\n"
static void FLIP(char *in_pos_line, char *out_pos_line) {
extended pos[3];
char temp[256];
extended ok=0.0;
asign(pos, 3, "", in_pos_line, "", -1000., 1000., ok, 1.); // asign copies eye position from string to pos array
// note: sometimes this has to accept missing values
if(pos[0] != MIS_VAL) {
pos[0] = - pos[0];
pos[2] = - pos[2]; // torsion
}
sprintf(temp, "%-6.2f %-6.2f %-6.2f ", pos[0], pos[1], pos[2]);
sprintf(out_pos_line, "pos: %s\n", temp);
}/*FLIP*/
void Execute_INN_POS(...){
...
...
...
switch(parameters.ruleType) {
case INN:
sprintf(inn_pos_string, PR_EXC_L, outExc[0], outExc[1], outExc[2], outExc[3], outExc[4], outExc[5]);
if( outPtr[0]) {
sprintf(outPtr[0], "inn: %s", inn_pos_string);
MovePtr(0);
} break;
case POS:
sprintf(inn_pos_string, PR_POS_L, rot_deg[0], rot_deg[1], rot_deg[2], trans[0], trans[1], trans[2]);
if( outPtr[0]) {
sprintf(outPtr[0], "pos: %s", inn_pos_string);
MovePtr(0);
}
break;
}
...
...
...
}/*Execute_INN_POS*/ The question in this case is, why does Orbit™ use the extended data type when it truncates intermediate calculation results (e.g. when flipping eye positions in the "FLIP" function above) to 2 digits after the comma. Another question is, why does function "FLIP" use 2 comma digits in its result when "Execute_INN_POS" uses 15 comma digits. In both cases double precision calculations would have been sufficient. We will not comment on performance issues and bad coding style, because we think that this speaks for itself. Interestingly, in the implementation of "FindPosition" and "FindInnervations" functions no rounding takes place and intermediate values are calculated with full extended precision.
- Orbit™ shows different convergence behavior when using consistent (extended) data types:
The convergence behavior of Orbit™ is highly fragile. Small changes in the source code (modification of precision as explained in (1), consistent usage of data types as explained in (2), consistent precision in rounding values as explained in (3)) immediately lead to different model predictions. Especially when the fixing eye is pathological and the torsional value of the fixing eye needs to be evaluated using the Newton optimizer, simulation results vary or loss of convergence results.
These problems clearly do not qualify Orbit™ to represent a solid basis for extensions or development of new models. Due to the fact that we cannot offer the source code of Orbit™ for download (for legal reasons) and only fragments of the source code would not help to illustrate this issue, it is not possible for us to give a short example.
- Orbit™ obviously moves "Pulleys" of superior and inferior obliques:
Up to our knowledge this is highly suspicious of being a bug, since the simulation of pulleys does not apply to the obliques. The default geometric configuration in Orbit™ for the superior oblique is to use the pulley as representation for the trochlea, whereas for the inferior oblique, origin and pulley are identical, since the inferior oblique does not have a pulley. Our example shows the Orbit™ left eye parameter editor as well as the mechanical stateviewer for a normal "healthy" left eye in primary position.
Under "Muscle Pulley Displacement", Orbit™ correctly shows 0 for both oblique muscles. However, under "Muscle Pulley Location", coordinates for inferior and superior obliques are different compared to the original geometrical default values shown in the left eye parameter editor. Consequently, there are two possibilities, either Orbit™ moves trochlea and non-existing pulleys (inferior oblique) but displays the wrong pulley displacement magnitudes, or Orbit™ does not move trochlea and inferior oblique pulley but coordinates are invalid. In either case, this is an inconsistency and speaking in Miller's words, "causes worry and doubt".
- 3D display (Live Eyes and Graphic Eyes) in Orbit™ incorrectly display insertion lines of eye muscles:
Since we did not review the 3D display source code, we insistently hope that insertion lines as seen in the 3D view are only for show and not part of the biomechanical model.
The animation above shows that obviously the insertion of a muscle in Orbit™ is not attached to the globe but rather "floats" on the globe. Looking at the red arrows in the animation, it can be seen that the distance between the cornea and the insertion line of the right medial rectus increases when the right eye is abducted, conversely for the left inferior rectus the distance to the cornea decreases in adduction of the left eye. Assuming that muscles are tightly coupled to the globe at their insertions, the situation shown in the animation above does not constitute a realistic simulation and causes worry and confusion.
- The support of Orbit™ for creating movies is simply ridiculous:
Creating movies in Orbit™ 1.8 is a time-consuming and difficult job for an ordinary computer user. The user interface support for creating movies is catastrophic and beyond any reasonable solution. In the User's Manual of Orbit™ 1.8, the following courageous explanation can be found under the topic "How do I make an Orbit Movie?":
"To use Movie Player to make a folder of PICS into a movie, the PICT files must all be named with the same characters followed by a positive number that tells Movie Player the order in which to string them together (eg: faden01, faden02, ...). You will probably need to rename the files produced by Orbit."
This cannot be seen as a feature of the program but rather looks like a complicated and clumsy workaround.
- Minor suggestion: Where is the "Mechanical Stateviewer for Inner Eye"? Even "incorrectly implemented" SEE++ has one!
- Minor GUI glitch:
A minor glitch in the textual display of the simulated deviations can be seen in the following image:
The first number of the horizontal deviation (exo-dev) is formatted differently. This is negligible, but compared to Miller's critique on SEE++ needs to be mentioned for the sake of completeness.
Differences between Orbit™ and the Orbit™ model in SEE++The above list shows the most important bugs we found in the Orbit™ 1.8 software system. In the following we want to give a short overview of the main differences between the Orbit™ 1.8 software system and the Orbit™ model in SEE++. In order to be able to compare simulation results of the Orbit™ model in SEE++ with the Orbit™ 1.8 software system running on Macintosh computers, we did NOT correct the bugs described in (1) and (5) in the Orbit™ model of SEE++.
The following list describes the most important differences:
-
SEE++ uses different signs for duction and torsion than Orbit™:
In the Orbit™ software system, adduction is defined as being negative, abduction is positive, intorsion is negative and extorsion is positive, respectively. In SEE++ (and also in the Orbit™ model of SEE++) the signs for duction and torsion are inverted, namely adduction is positive, abduction is negative, intorsion is positive and extorsion is negative. Since the convention of signs does not have any influence on simulation results we decided to use a common clinical convention which defines positive angles towards the nose for all three degrees of freedom.
- SEE++ uses Quaternions to represent rotations and eye positions:
Orbit™ uses rotation matrices and radian angles to represent rotations and eye positions. SEE++ uses Quaternions which have some advantages over other representations of rotations:
- Quaternions don't suffer from gimbal lock, unlike euler angles.
- They can be represented as 4 numbers, in contrast to the 9 numbers of a rotations matrix.
- The conversion to and from axis/angle representation is trivial.
- Smooth interpolation between two quaternions is easy (in contrast to axis/angle or rotation matrices).
- After a lot of calculations on quaternions and matrices, rounding errors accumulate, so quaternions need to be normalized and rotation matrices need to be orthogonalised, but normalising a quaternion is a lot less troublesome than orthogonalising a matrix.
- Similar to rotation matrices, you can just multiply 2 quaternions together to receive a quaternion that represents both rotations.
-
Orbit™ gaze selector always shows points in the coordinate system of the following eye:
The SEE++ system contains a so-called "gaze pattern dialog" as an equivalent to the Orbit™ gaze selector. However, in Orbit™ the points selected in this dialog are always in the coordinate system of the following eye, whereas in SEE++ points used as intended gaze positions for the simulation are defined in the coordinate system of the fixing eye (with a different sign in duction). Moreover, the gaze pattern dialog in SEE++ can also be used to enter patient-measured data, which are additionally displayed in Hess diagrams and defined in the coordinate system of the following eye. Orbit™ uses a "measured gaze editor" for entering patient-measured data, which is used in a completely different way than the Orbit™ gaze selector. Since we think it is a bad idea to enter intended and measured gaze positions in a completely different manner, SEE++ uses the same input procedure for both types of gaze positions.
- SEE++ consistently uses double precision for floating point calculations:
As already explained in (2) and (4) in the above list, Orbit™ uses the "extended" data type for most calculations. Since SEE++ is designed to be run on Intel platforms (Microsoft Windows operating system) where no such data type exists, SEE++ consistently uses double precision for all floating point calculations. The "extended" data type on Macintosh is somewhat historic, comes from the old Mac generation and in case this type is used in implementations, unfortunately makes it impossible to identically port software between platforms.
Most modern systems handle floating point following the IEEE-695 standard. However, there are still portability issues. Most processors use 64 bits of precision when computing floating point values. However, the widely used Intel x86 series of processors compute temporary values using 80 bits of precision, as do most instances of the Motorola 68k series. Some other processors, such as the PowerPC, provide fused multiply-add instructions which perform a multiplication and an addition using high precision for the intermediate value. Optimizing compilers will generate such instructions based on sequences of C/C++ operations.
For almost all programs, these differences do not matter. However, for programs which do intensive floating point operations, the differences can be significant. It is possible to write floating point loops which terminate on one sort of processor but not on another.
Unfortunately, there is no rule of thumb that can be used to avoid these problems. Most compilers provide an option to disable the use of extended precision for intermediate values (for GNU cc, the option is `-ffloat-store'). However, this merely shifts the portability problem elsewhere.
When comparing simulation results between Orbit™ 1.8 Macintosh version and the SEE++ implementation of Orbit™, differences only occur when numerical optimization for some gaze position reaches the precision limit of double floating point precision. In this case, either the resulting gaze position will show a small difference (we observed differences up to 0.1 degrees), or in very special cases, loss of convergence occurs and SEE++ will not find a solution for this gaze position.
Another portability problem is the calculation of trigonometric functions, usually provided by the runtime environment that comes with each compiler on each operating system. Due to different architectures, it cannot be guaranteed that runtime functions behave identically, of course this will never be the case when "extended" data type is used on Mac platforms.
- SEE++ uses different "normal" eye parameters than Orbit™:
In SEE++ "normal" eye parameter values have been used according to Volkmann and Miller, however Miller revised his normal eye parameter values in Orbit™ 1.8 again. Differences are rather marginal, but when comparing simulation results, normal eye values must certainly be the same for both programs. SEE++ therefore includes default scenarios for normal eye data as "Volkmann Data" (no pulleys), "Pulley Data" (default values from Volkmann and Miller) and "Orbit Data" (latest version of Miller's "Orbit_Norm" file).
In the picture below these default scenarios are shown and the respective scenario needs to be selected in order to load the correct normal eye data prior to simulation and comparison of results.

- Comparing simulation results between Orbit™ 1.8 Macintosh and Orbit™ implementations in SEE++:
We would like to briefly show a comparison of a simulation between the two programs Orbit™ 1.8 and Orbit™ implementation within SEE++. For simulation we modified distribution of innervations of the abducens nerve to innervate medial and lateral rectus muscles of a left eye with a magnitude of 60%.
Below the parameter values that were changed for this simulation are shown in Orbit™ and SEE++:

- The simulation results show that the calculated eye positions are exactly the same in both programs (see textual deviation angles), however, SEE++ omits one fixation point due to loss of convergence explained in (4). This is the rare case, more frequently deviation angles will show up to 0.1 degrees of difference between both programs.

- In order to verify that both implementations are identical, stateviewer values can be compared, and it can be seen that all calculated model values for eye positions are the same. Below the comparison of the fixation of the left eye at 30 degrees abduction is shown, consequently the right eye will move into adduction. Please also note that duction and torsion have opposite signs in SEE++ compared to Orbit™.

- Innervations in both programs show identical values, however, in SEE++ medial and lateral rectus muscles are swapped:

- Muscle forces are almost identical, inferior oblique total longitudinal muscle force differs (0.1g difference) due to floating point inconsistencies as explained in (4):

- Muscle geometry shows identical values for all parameters in both programs:

- In case of sceptisism, we encourage the reader to download both programs and gather own experiences (if one can find a Macintosh that is old enough to run Orbit™ 1.8). Unfortunately, as requested by Miller, it is impossible to specify in which situations differences occur due to the nature of floating point arithmetic and processor architectures. However, in all our numerous experiments, we never found differences greater that 0.1 degrees in resulting Hess-Lancaster simulation results.
In case that some reader reached this point of this document, we would be happy to hear your opinion and would like to receive feedback to